Interactions and heating time-course of cerebral aneurysm flow diverters during scanning at 7 Tesla

Suguru Igarashi1, Kohei Chida1, Tsuyoshi Matsuda2, Kota Takeda2, Susumu Hayakawa2, Shota Tsutsui1, Shunrou Fujiwara1, Yoshitaka Kubo1, Makoto Sasaki2, and Kuniaki Ogasawara1
1Department of Neurosurgery, Iwate Medical University, Morioka, Japan, 2Division of Ultrahigh Field MRI, Institute for Biomedical Sciences, Iwate Medical University, Morioka, Japan

Synopsis

Keywords: Interventional Devices, Stroke, flow diverter, 7T MRI

Motivation: Adverse interactions with static magnetic fields (e.g., displacement force and torque) and radiofrequency-induced heating during MRI have been resolved for 3T MR systems, but have yet to be assessed for 7T MRI.

Goal(s): The present study aimed to assess displacement force and torque in the 7T static magnetic field and to clarify radiofrequency-induced heating during 7T MRI for two types of FDs.

Approach: This study was conducted based on the ASTM standards.

Results: Magnetic field interactions and heating on FDs during 7T MRI are acceptable from a safety perspective.

Impact: This study based on ASTM standards showed magnetic field interactions and heating on FDs during 7T MRI are acceptable from a safety perspective.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords