Keywords: Interventional Devices, Stroke, flow diverter, 7T MRI
Motivation: Adverse interactions with static magnetic fields (e.g., displacement force and torque) and radiofrequency-induced heating during MRI have been resolved for 3T MR systems, but have yet to be assessed for 7T MRI.
Goal(s): The present study aimed to assess displacement force and torque in the 7T static magnetic field and to clarify radiofrequency-induced heating during 7T MRI for two types of FDs.
Approach: This study was conducted based on the ASTM standards.
Results: Magnetic field interactions and heating on FDs during 7T MRI are acceptable from a safety perspective.
Impact: This study based on ASTM standards showed magnetic field interactions and heating on FDs during 7T MRI are acceptable from a safety perspective.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords