An elliptic-bore low-field portable Halbach magnet with full-length gradient support

Fernando Galve1, Eduardo Pallás1, Teresa Guallart-Naval1, Pablo García-Cristóbal1, José Miguel Algarín1, and Rubén Bosch1
1Institute for Molecular Imaging and Instrumentation (CSIC-UPV), Valencia, Spain

Synopsis

Keywords: Low-Field MRI, Low-Field MRI, Magnets (B0), New Devices

Motivation: To build a truly portable MRI scanner for neurological and MSK applications, optimized for field homogeneity and FoV size compared to standard Halbach configurations.

Goal(s): To explore elliptic-bore magnets to improve the field homogeneity and remove constraints to the extent of the gradient coils typical of Halbach magnets.

Approach: We have optimized a tightly-packed distribution of magnetic Nd2Fe14B cubes with differential evolution algorithm, and a second array of shimming magnets with interior point and differential evolution methods.

Results: We have built a 10 & 14 cm radii elliptic magnet with 87.4 mT@5700 ppm in a 10 cm radius FoV.

Impact: Elliptic-bore Halbach magnets enhance the ergonomicity and field distribution of low-costportable MRI scanners, while allowing for full-length gradient support to increase the FoV. Thisgeometry can be potentially adapted for a prospective low-cost whole-body technology.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords