Meeting Banner
Abstract #3193

Development of a transmit gain calibration method using the Bloch-Siegert shift for nonproton hyperpolarized MRI on a preclinical 7T scanner

Collin J. Harlan1,2, Keith A. Michel2,3, and James A. Bankson1,2
1The University of Texas MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, TX, United States, 2Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, United States, 3Imaging Physics Residency Program, The University of Texas MD Anderson Cancer Center, Houston, TX, United States

Synopsis

Keywords: Hyperpolarized MR (Gas), Hyperpolarized MR (Gas), Transmit Gain Calibration, Bloch-Siegert Shift

Motivation: Transmit gain calibration is necessary for nonproton hyperpolarized MRI to accurately calibrate excitation angles.

Goal(s): To implement a phase-based TG calibration method using the Bloch-Siegert shift for nonproton HP MRI on a preclinical 7T scanner.

Approach: An off-resonance Fermi-shaped Bloch-Siegert pulse follows directly an excitation pulse. Percent difference between our measured Bloch-Siegert TG calibration factors and the validation TG calibration factors for 1H and 13C were calculated.

Results: Our method allows for fast, accurate, and robust TG calibration of preclinical nonproton HP MRI studies with minimal error (<5%) for 1H and 13C.

Impact: This method will facilitate efficient transmit calibration for 13C and HP 129Xe MRI on a preclinical 7T scanner, where limited bore size and the inability to easily implement a thermal calibration phantom can make 129Xe TG calibration a challenge.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords