Meeting Banner
Abstract #4554

Comparative Analysis of Cardiac T1 Mapping Reliability: Free-Breathing Versus Breath-Hold Techniques

Julia V Velikina1, Orhan Unal1, and Alexey A Samsonov1
1University of Wisconsin - Madison, Madison, WI, United States

Synopsis

Keywords: Quantitative Imaging, Myocardium, T1 Mapping, Repeatability

Motivation: Quantitative myocardial T1 mapping is an invaluable tool for detection and characterization of myocardial fibrosis and infarction. Recent emergence of free-breathing non-ECG-gated techniques requires evaluation of their reliability in controlled experiments.

Goal(s): We evaluated repeatability of free-breathing non-ECG-gated Multitasking and breath-held MOLLI T1 mapping techniques and studied factors influencing them.

Approach: Test-retest study was performed in 15 healthy adult volunteers. Bland-Altman plots and Spearman rank correlation were used for statistical analysis.

Results: Our analysis reveals that while free-breathing technique has low bias, controlling data sampling patterns is needed to prevent high variability of test-retest measurements.

Impact: Reliability of myocardial T1 mapping with free-breathing techniques may benefit from designing controlled data sampling patterns utilizing either external or self-calibrated cardiac/respiratory triggers.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords