3DMRCP Quantitative Image Quality: A Comparison Between CS-SENSE And SENSE And Its Acquisition Time

Chee Chian Teo1
1Diagnostic Radiology, Tan Tock Seng Hospital, Singapore, Singapore

Synopsis

Keywords:

Motivation: Upgrading of MR scanner with compressed sensing (CS) could reduce acquisition time. Image quality of 3DMRCP with CS, however, need to be assessed quantitatively to ensure its standard despite of acceleration.

Goal(s): We aimed to prove that 3DMRCP with CS can be employed in clinical practice with comparable quantitative image quality to SENSE and with shorter acquisition time.

Approach: Quantitative image quality (signal intensity, noise, SNR and CNR) of 3DMRCP with CS and SENSE of the same subject that acquired at two different timepoints were compared retrospectively.

Results: 3DMRCP with CS have maintained the quantitative image quality and has a shorter scan duration.

Impact: 3DMRCP with CS only not can achieve higher spatial resolution but also with a shorter acquisition time while maintaining its quantitative image quality. Therefore, 3DMRCP with CS is recommended to be adopted in daily clinical practice.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords