Meeting Banner
Abstract #5159

Study on 2 Channel Application of Microscopy Coil Using Self-made Phantom

Han Byul Chung1, Young Mi Kim1, Eun Hee Seo1, and Chang Min Dae1
1Radiology, Seoul National University Bundang Hospital, Seongnam-si, Korea, Republic of

Synopsis

Motivation: Microscopy coil has high spatial resolution and SNR, but it has a limited effective Signal Intensity(SI) range.

Goal(s): This study suggests a method to improve SI & effective SI range by various combinations of Microscopy Coil.

Approach: We made a MRI Marker Phantom and analyzed SI using Microscopy Coil with 1channel (23mm, 47mm), 2channel (23+47mm, 23+23mm, 47+47mm). We measured effective SI range and SI.

Results: In the results, SI of 2channel was higher than that of 1channel in all 12 ROI. Furthermore, change of changing of sense factor showed no significant change. Effective SI range was 47+47mm, 23+47mm, 47mm, 23+23mm, 23mm coil respectively.

Impact: Various combinations of Microscopy 2channel Coil had advantage of increasing Signal Intensity(SI) and effective SI range compared to 1channel. Also use of sense factor in 2channel can reduce scan duration without SI change for patients who have difficulty in cooperating.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords