An Experimental Comparison for RF Heating of DBS Electrodes at 3T and 7T

Nur Izzati Huda Zulkarnain-Lemke1, Alireza Sadeghi-Tarakameh1, Jeromy Thotland1, Noam Harel1, and Yigitcan Eryaman1
1Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, MN, United States

Synopsis

Keywords: Safety, Safety

Motivation: With the growing demand for MRI scans in patients with deep brain stimulation (DBS) implants, understanding SAR and RF heating around these devices is essential for safety and improved scan protocols.

Goal(s): To compare the RF heating around DBS electrodes at 3T and 7T.

Approach: Bilateral DBS were fixed in the skull cavity of an anthropomorphic head phantom. Temperature-based SAR estimation quasi-static electric field and thermal simulations were conducted at 7T and 3T.

Results: The mean of the 1 g-averaged SAR normalized to B1+RMS2at 7T, on average, was 5.6 times lower than at 3T.

Impact: Comparing 1 g-averaged SAR (normalized to B1+RMS2) around DBS electrodes at 3T and 7T suggested that 7T MRI may pose less heating risk and provided a compelling case to expand the potential of imaging DBS patients at higher-field MRI.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords