Keywords: Data Processing, Data Processing, Low‐Field MRI, lung, Motion Correction
Motivation: Current gold standard for lung imaging is computed tomography (CT). This method only provides morphologic data and utilizes ionized radiation. MR imaging has not reached clinical applications at traditional field strengths. At low-field strengths MRI may be uniquely beneficial to the imaging of the lung parenchyma and function
Goal(s): The creation of a 3D cine protocol to study the functional dynamics of the lung on a low-field 0.55T MR.
Approach: We developed a free-breathing pipeline to create 3D cine images from 2D slices.
Results: We were able to reconstruct 12 volumes along a respiratory cycle with a total acquisition time of ~15 minutes.
Impact: Low-field 0.55T MRI enables functional lung evaluation without ionizing radiation, providing a safer alternative to traditional CT scans. This approach might improve the assessment of lung pathologies, offering enhanced diagnostic and monitoring capabilities, particularly for functional lung imaging.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.