Keywords: Parallel Transmit & Multiband, Parallel Transmit & Multiband
Motivation: At 7T, slice-by-slice 2D turbo-spin-echo (TSE/RARE/FSE) images are needed for clinical protocols, but show strong signal drop-outs in single transmit scans.
Goal(s): Our goal is to acquire high-fidelity 2D TSE images across the whole brain using parallel transmit.
Approach: PTx pulses must satisfy the Carr-Purcell-Meiboom-Gill (CPMG) conditions for good image quality. We achieve this using novel slice-selective scalable spokes pTx pulses.
Results: Simulations and phantom scans show that 3-spoke scalable pulses have high flip-angle uniformity and maintain consistent phase profiles. In vivo tests show uniform images across the whole brain including in temporal lobes. VOP SAR supervision is needed for optimum TR.
Impact: Scalable spokes pTx pulses produce high quality 2D-TSE images. It mitigates signal dropouts and recover diagnostic quality when single-transmit scans are inadequate. This will be valuable for patients with drug-resistant focal epilepsy to find epilepsy-causing lesions in the temporal lobes.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.