Implant-Friendly RF Excitation Solutions for Patients with Bilateral DBS Electrodes

Nur Izzati Huda Zulkarnain-Lemke1, Alireza Sadeghi-Tarakameh1, Jeromy Thotland1, Noam Harel1, and Yigitcan Eryaman1
1Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, MN, United States

Synopsis

Keywords: Safety, Safety

Motivation: To develop safe MRI procedures for patients with deep brain stimulation (DBS).

Goal(s): To investigate the feasibility of an implant-friendly optimization as a safety pre-scan workflow.

Approach: The RF induced currents on DBS electrodes were measured using an image based method. The two-channel RF excitation patterns were optimized to minimize heating and RF artifacts around bilateral DBS electrodes in three configurations.

Results: The optimization reduced induced current and RF artifacts near the electrodes. The pre-scan procedure can be completed in less than 6 minutes of active scanning time.

Impact: We demonstrated a fast, patient-specific optimization workflow to improve the safety of MRI scans for DBS patients. The workflow is efficient enough to be seamlessly integrated into routine imaging procedures before the main study scan.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords