An Automated Validation System for Active Implantable Medical Device (AIMD) Model under MR RF Field Exposure

Ziyu Zuo1, Ao Shen1, Farshad Ebrahimi1, Qingyan Wang1, Hongbae Jeong2, Ananda Kumar2, and Ji Chen1
1University of Houston, Houston, TX, United States, 2US Food and Drug Administration, Silver Spring, MD, United States

Synopsis

Keywords: Hybrid & Novel Systems Technology, Safety, Active Device

Motivation: Current AIMD model validation method is expensive, time-consuming, and subject to human operational error. An automated robotic system for testing would streamline validation testing.

Goal(s): To develop a novel automated AIMD model validation system that is cost-effective and accurate.

Approach: A novel system is developed by integrating the a robotic arm, a box-shaped phantom, a signal generator and power amplifier, and a fiber optic temperature measurement apparatus to streamline the validation.

Results: An automated validation system for both 1.5T and 3T is constructed with minimum human error.

Impact: A cost-effective and yet more accurate AIMD model validation system/method is developed following ISO 10974 standard. Compared to the traditional birdcage coil validation method, this system requires less labor, lower cost, and has lower systematic error.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords