Motivation: To demonstrate whether SmartSpeed Motion Free an alternative to Cartesian scan and MultiVane XD in terms of basic image characteristics.

Goal(s): To compare the basic image characteristics of the Cartesian scan, MultiVane XD, and SmartSpeed Motion Free in the absence of patient motion.

Approach: Contrast ratio, contrast to noise ratio, noise power spectrum, and contrast response for the sharpness were evaluated in the phantom experiments.

Results: SmartSpeed Motion Free has the same contrast as Cartesian scan and MultiVane XD, with less noise and a frequency characteristic similar to the Cartesian scan. Furthermore, it had better sharpness than MultiVane XD.

Impact: SmartSpeed Motion Free has the potential to provide superior image quality compared to MultiVane XD and image quality closer to Cartesian scan in the same scan time, even if patient motion has not occurred.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords