Sonal Josan1, Andrew B. Holbrook, 12,
Elena Kaye, 1,3, Christine Law, 1,3, Kim Butts Pauly1
1Radiology, Stanford University,
Stanford, CA, United States; 2Bioengineering, Stanford University,
Stanford, CA, United States; 3Electrical Engineering, Stanford
University, Stanford, CA, United States
MR
thermometry relies on the proton resonance frequency shift with temperature,
which can produce off-resonance artifacts in EPI and spiral sequences. This
work analyzed the appearance of the focused ultrasound (FUS) hotspot on
several EPI and Spiral trajectory designs through simulations and FUS
experiments. The distortion of the FUS spot with single shot EPI or Spiral
imaging can be severe for the high temperature changes used in ablation, and
may lead to under-estimation of the peak temperature. Multi-shot sequences
can be used to reduce the shifts/distortion to a tolerable level.