A
  method that allows myocardial first-pass perfusion measurements in mice is
  presented. Using a combination of segmented saturation-prepared FISP
  acquisition and GRAPPA parallel imaging allows a temporal resolution of one
  image every three heart beats with an acquisition time of less than 16 ms.
  First-pass perfusion images showed the influx of contrast agent into the
  myocardium with sufficient temporal resolution to derive semi-quantitative
  perfusion values. These were found significantly lower in a mouse with
  myocardial infarction compared to healthy control mice.