Abstract #1696
Uncertainty Quantification of Multi-Site T1 Measurements with Polyvinylpyrrolidone (PVP) Phantom and Human Brain Using Wild Bootstrap Analysis
Congyu Liao 1 , Meng Chen 1 , Darong Zhu 2 , Hongjian He 1 , Song Chen 1 , Qiuping Ding 1 , and Jianhui Zhong 1
1
Center for Brain Imaging Science and
Technology, Zhejiang University, Hangzhou, Zhejiang,
China,
2
Hangzhou
First People's Hospital, Zhejiang, China
In this study, T1 mapping of a phantom with different
concentration of Polyvinylpyrrolidone (PVP) solutions
and a volunteer were acquired in two different 3T
scanners using DESPOT1 method, and wild bootstrap
analysis was used to assess both intra- and inter-site
uncertainty of quantitative T1 measurements.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.