Inter-scanner reproducibility of 4 minute whole brain myelin mapping using FAST-T2

Thanh D. Nguyen1, Kofi Deh1, Sneha Pandya1, Yi Wang1, and Susan A. Gauthier1

1Weill Cornell Medical College, New York, NY, United States

The purpose of this study was to measure the inter-scanner reproducibility of fast myelin water fraction (MWF) mapping using a 4 min FAST-T2 sequence on three scanners from two vendors at 1.5T and 3T. Regional MWF measurements obtained in 7 healthy volunteers were highly reproducible with excellent inter-scanner correlation (R>0.95) and small MWF bias of less than 1%.