More and faster: multi-timepoint ASL at 150ms time-resolution with whole brain coverage by combining time-encoding, Look-Locker, Multi-Band and flip-angle sweep

Merlijn C.E. van der Plas1, Wouter M. Teeuwisse1, Sophie Schmid1, and Matthias J. van Osch1

1C.J. Gorter Center for high field MRI, Department of Radiology, Leiden University Medical Center, Leiden, Netherlands

Multi-timepoint ASL can be acquired with several methods; however, most of these methods have some confounding aspects, such as loss of temporal resolution or limited brain coverage. We combine Hadamard-8 time-encoded-pCASL with Look-Locker readout (4 phases×150ms) for a high temporal resolution; moreover, multiband-factor 4 is used for a time efficient acquisition manner enabling whole brain coverage. This combination of techniques results in 25 PLDs for 16 slices, which provides information about the inflow of labeled blood and tissue perfusion. These images have a temporal resolution of 150ms, whereas inflow of label is measured at 75ms resolution.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords