Accuracy of ADC measurements with an Ultrashort Echo Time Diffusion Weighted stimulated echo 3D Cones sequence (DW-STEAM 3D Cones UTE)

Paul Baron1, Dirk H.J. Poot1,2,3, Piotr A. Wielopolski1, Edwin H.G. Oei1, and Juan A. Hernandez-Tamames1

1Radiology and Nuclear Medicine, Erasmus Medical Center, Rotterdam, Netherlands, 2Medical Informatics and Radiology, Erasmus Medical Center, Rotterdam, Netherlands, 3Imaging Science and Technology, Delft University of Technology, Delft, Netherlands

Diffusion weighted STEAM with UTE readout is a promising acquisition method to measure diffusion in tissues with short T2 and T2*, such as tendons, ligaments and menisci. However, the accuracy of the ADC obtained with this method has not been studied before. Using experiments and Bloch simulations we show that the ADC can be biased, especially when a short TR is used, and that this bias depends on T1 and T2. Randomization of the diffusion gradient direction reduces the bias, providing clear suggestions to improve acquisition and/or post processing that reduces the ADC bias.