A Feedback System for Molecular Imaging of 19F FDG

Mirko Hrovat1, James Balschi2, Mike Dabaghyan1, Mark Mattingly3, and Neal Kalechofsky2

1Mirtech, Inc., Boston, MA, United States, 2Millikelvin Technologies, LLC, Braintree, MA, United States, 3Bruker BioSpin Corp., Billerica, MA, United States

A feedback system used in conjunction with a high concentration spin reservoir demonstrates an enhanced MRI signal from a molecular imaging target at low concentration. Enhanced images of a 10mM 19F FDG target obtained by a modified 3D Flash sequence are shown. Detectability curves reveal that for a 100ul voxel size, an imaging time of ~30 min is required to image 1mM 19F FDG with an SNR of 4.