Meeting Banner
Abstract #5047

Bloch-Siegert Phase-Encoded MRI with a Single RF Coil and Frequency-Swept Pulses

Christopher J Hasselwander1 and William A Grissom1

1Biomedical Engineering, Vanderbilt University, Nashville, TN, United States

RF encoding using the Bloch-Siegert shift has the potential to replace conventional gradient encoding with cheaper RF gradients, and can directly replace gradient pulsing in common MRI sequences. However, current implementations require large frequency offsets to prevent on-resonance excitation, which subsequently requires high RF power for encoding as well as separate imaging and encoding RF coils. Here we show that frequency-swept encoding RF pulses enable the use of a single RF coil for imaging and encoding since the frequency offset can be brought much closer to resonance.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords