Homogenous Excitation in Whole Brain CEST: Combination of Snapshot CEST and Multiple Interleaved Mode Saturation

Andrzej Liebert1, Moritz Zaiss2, Rene Gumbrecht3, Patrick Liebig3, Benjamin Schmitt3, Frederik Laun1, Arnd Doerfler4, Michael Uder1, and Armin M. Nagel1

1Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 2Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 3Siemens Healthcare GmbH, Erlangen, Germany, 4Department of Neuroradiolohy, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany

To perform Chemical Exchange Saturation Transfer MRI of the whole brain a homogeneous saturation and fast readout are required. To achieve a fast and robust 3D acquisition a spiral-centric-reordered GRE readout was used. In addition, a Multiple Interleaved Mode Saturation scheme was applied to mitigate B1+-inhomogeneity effects of the CEST saturation. Combination of these two methods allows acquiring a homogenous CEST contrast in a volume of approximately 220x220x45mm3 within an acquisition time of 7 min 26s.