With the number of 7T sites increasing, more brain-MRS studies will be conducted at 7T. Particularly GABA and GSH are of interest and benefit from the increased resolution. To date, studies have applied both conventional and edited MRS to measure these metabolites, however, there is no consensus on whether to use edited MRS. This study shows that it depends on brain region and potentially voxel size and number of acquisitions, whether it is preferable to use editing for GABA. There is less reason to argue that editing benefits measurement of GSH, but this study indicates a slight preference for editing.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords