Water/fat separation can be performed by exploiting chemical shift, which is proportional to field strength. Hence, on lower-field MRI systems (0.1 … 1T), the absolute resonance frequency difference is reduced compared to clinical high-field MRI, raising the question if water/fat separation can still be reliably performed despite the proximity of the resonances. In this work, we examined the feasibility of Dixon-type water/fat separation in the abdomen on a 0.75T lower-field MRI system. We show that Cartesian multi-acquisition and multi-echo as well as spiral multi-acquisition sequences provide reliable water/fat separation despite the reduced water/fat resonance shift of approximately 108 Hz.