Frequency sweep 31P MR spectroscopic imaging at 7T

Songi Lim1,2, Mark Stephan Widmaier1,2, and Lijing Xin1,3
1CIBM Center for Biomedical Imaging, Lausanne, Switzerland, 2Laboratory for Functional and Metabolic Imaging, EPFL, Lausanne, Switzerland, 3Animal Imaging and Technology, EPFL, Lausanne, Switzerland

Synopsis

Phase-cycled spectroscopic imaging (PCSI) method was implemented and validated for 31P PCSI imaging at 7T. The PCSI method uses a balanced steady-state free precession sequence with an ultra-low flip angle (<1°) to achieve sharp passband with 2.52-ms of TR, which enable to accelerate the acquisition. With prior knowledge of 31P spectra, it is feasible to acquire major 31P peaks by changing the frequency offset and non-uniform phase sweeping instead of acquiring full spectra uniformly. To investigate feasibility of the method, a multi-compartment KH2PO4 phantom with in vivo equivalent concentrations was prepared and 31P PCSI was compared with conventional FID-CSI.