Synopsis

Keywords: fMRI Analysis, fMRI, NORDIC , VASO, laminar fMRI

Motivation: NORDIC denoising can effectively enhance the limited SNR in high-resolution fMRI. However, its application on VASO is yet to be validated.

Goal(s): We aimed to evaluate applications of NORDIC on VASO data and to offer recommendations for its execution.

Approach: We examined NORDIC’s capability to suppress noise while preserving the VASO signal across a wide parameter spectrum

Results: With a proper set of parameters, NORDIC effectively suppressed noise with minimal biases on the underlying signal.

Impact: NORDIC can substantially enhance the SNR in submillimeter VASO fMRI. We found the denoising performance to be sensitive to parameter choices and provide recommendations for safe execution.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords