Initial clinical evaluation of a 60-channel small loop receive coil array for high-resolution pediatric MRI

Marine Moeremans1, Wonje Lee1, Praveen Jayapal1, Fraser Robb2, John Pauly1, Shreyas Vasanawala1, and Ali B. Syed1
1Stanford University, Stanford, CA, United States, 2Ge healthcare coils, Ohio, OH, United States

Synopsis

Keywords: Body, Pediatric

Motivation: Commercial MRI coils typically offer suboptimal parallel imaging and SNR performance for young children.

Goal(s): Construct & evaluate the clinical feasibility of using a highly flexible receive array with small coil loops to image the pediatric chest, abdomen, and pelvis.

Approach: Nine patients were scanned with the investigational coil and 9 age- and anatomy-matched control patients were identified. Qualitative grading of image quality metrics was performed by two radiologists.

Results: The investigational coil achieves similar image quality to standard coils. Protocol modification to leverage optimal coil geometry is needed.

Impact: This work provides initial clinical feasibility on using a small-loop receive coil array for imaging of the chest, abdomen, and pelvis. Diagnostic image quality was obtained in all cases, indicating the small element size is feasible in clinical use.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords