Keywords: CEST / APT / NOE, CEST & MT
Motivation: Steady state CEST (ssCEST) is an approach to accelerate CEST experiments, which typically require long measurement times. The ssCEST-signal generally depends not only on the concentration and exchange rate of the exchanging protons but also on the T1 relaxation of the tissue.
Goal(s): The aim of the work was to develop a T1-independent ssCEST method.
Approach: Based on the analytical equation for spoiled gradient echo ssCEST-signals, a new T1-independent evaluation metric based on a double-angle approach was derived and tested using simulations and phantom measurements.
Results: The T1-independence of the derived evaluation metric and its applicability to potatoes as plant phantoms was demonstrated.
Impact: The T1-dependence of spoiled gradient echo ssCEST-signals can be eliminated and corrected by using the proposed analytical evaluation metric based on a double-angle approach.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.