The optimized double echo steady-state (DESS) sequence for T2w fluid attenuation contrast at 7T

Qing Li1, Ruiting Chen2, Dongcui Wang2, Yicheng Hsu1, Jianxun Qu1, and Weihua Liao2
1MR Research Collaboration Team, Siemens Healthineers Ltd., Shanghai, China, 2Department of Radiology, Xiangya Hospital, Central South University, Shanghai, China

Synopsis

Keywords: Signal Modeling, High-Field MRI, DESS

Motivation: Three-dimensional high-resolution fluid attenuation inversion recovery (FLAIR) acquired using the SPACE sequence at 7T is challenging due to the SAR issue and B0 inhomogeneity.

Goal(s): To optimize the DESS sequence for presenting a semi-FLAIR contrast

Approach: The effect of the excitation FA and the spoiler gradient between the first (S_plus) echo and the second (S_minus) echo of the DESS was evaluated by EPG simulation and validated by the in vivo experiments.

Results: The optimized DESS sequence (low FA of 8° and spoiler gradient) produces perfect fluid attenuation near the sphenoid sinus with a scan time of 6:30 min (0.64 mm iso).

Impact: This study reported the first time the semi-FLAIR contrast at 7T without SAR limitation and B0 impact especially around sphenoid sinus using the low flip angle DESS sequence.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords