Keywords: Hyperpolarized MR (Gas), Hyperpolarized MR (Gas), xenon-129
Motivation: Rapid hyperpolarized xenon-129 production is essential for hyperpolarized gas MRI studies.
Goal(s): To improve hyperpolarized xenon-129 production rates, for rapid on-demand dose production.
Approach: A continuous-flow polarizer was optimized for maximum xenon-129 polarization and cryogenic volume recovery at high xenon flow rate (180 ml/min).
Results: Maximum xenon-129 polarization of 10.2 ± 1.3 % was measured at 3 bar cell pressure using 3% Xe (86% enriched xenon-129), 10% N2 and 87% He gas mixture. Cryogenic Xe volume recovery was improved by using high cell pressure and a large cold finger. 1 liter of hyperpolarized xenon-129 was produced in 6.1 minutes, to polarization of 9.98%.
Impact: The results demonstrate the ability to produce rapid on-demand hyperpolarized xenon-129 doses in large volumes, which will improve clinical workflows for hyperpolarized gas MRI studies.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.